Quantitative determination of total cesium in highly active liquid waste by using liquid electrode plasma optical emission spectrometry

Do, V. K.; Yamamoto, Masahiko; Taguchi, Shigeo; Takamura, Yuzuru*; Surugaya, Naoki; Kuno, Takehiko

Talanta, 183, p.283 - 289, 2018/06

 Times Cited Count:10 Percentile:39.09(Chemistry, Analytical)

We develop a novel analytical method employing liquid electrode plasma optical emission spectrometry for measurement of total cesium in highly active liquid wastes. Limit of detection and limit of quantification are 0.005 mg/L and 0.02 mg/L, respectively. The method is validated and applied to the real samples.