Application of a portable pulsed magnet to synchrotron radiation experiments

Matsuda, Yasuhiro*; Ueda, Yuji*; Nojiri, Hiroyuki*; Takahashi, Toshiharu*; Inami, Toshiya; Owada, Kenji; Murakami, Yoichi; Arima, Takahisa*

A very small capacitor bank and miniature pulsed magnets have been developed to perform high-field synchrotron radiation experiments. A 20T pulsed field can be generated using an energy of 1kJ with typical pulse duration about 1ms. Electron spin resonance experiments in millimeter wave range and high-field X-ray diffraction experiments have been done successfully.