Determination of technetium-99 and neptunium-237 in environmental samples by inductively coupled plasma mass spectrometry

Sumiya, Shuichi ; Morita, Shigemitsu; Tobita, Kazunori; Kurabayashi, Mizumi

New analytical techniques using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) were applied to the determination of technetium-99 (Tc-99) and neptunium-237 (Np-237) in environmental samples. The determination of Tc-99 consists of a cyclohexanone solvent extraction method to eliminate the interference isobar (Ru'99). An anion exchange with acetic acid media and a ITA-xylene solvent extraction were used for the determination of Np-237. Technetium-95m and Np-239 were used as chemical recovery tracers in these methods. The concentrations of Tc-99 and Np-237 in the typical surface soil samples in Japan ranged form 8.1$$times$$10$$^{-2}$$ to 1.8$$times$$10$$^{-1}$$ Bq/kg-dry, from 3.3$$times$$10$$^{-3}$$ to 8.0$$times$$10$$^{-3}$$ Bq/kg-dry, respectively.