Consideration on calibration and correction factors of an $$H$$$$_{rm P}$$(10) chamber for different radiation qualities and angles of incidence

Mikami, Satoshi   ; Iti$'e$, C.*; Texier, C.*

This paper presents the characterization, performed in IRSN (France) of a $$H$$$$_{rm P}$$(10) chamber in terms of calibration factor and correction factors for the radiation qualities of ISO narrow spectrum series. The $$H$$$$_{rm P}$$(10) chamber can measure directly the conventional true value of Hp(10) after a calibration by a reference laboratory, and can by used for transferring reference quantities $$H$$$$_{rm P}$$(10) from a reference laboratory.