Response of a high-resolution high-energy photon spectrometer (HHS) to monochromatic high-energy $$gamma$$ rays

Harada, Hideo   ; Furutaka, Kazuyoshi  ; Nakamura, Shoji  ; Osaka, Koji*; Akimune, Hidetoshi*; Utsunomiya, Hiroaki*; Osaki, Toshiro*; Igashira, Masayuki*

The response function of a high-resolution high-energy photon spectrometer (HHS), consisting of a twin Ge detector and BGO (Bi$$_{3}$$Ge$$_{4}$$O$$_{12}$$) anti-coincidence shields, has been measured to 10.763 MeV monochromatic $$gamma$$-rays generated via an$$^{27}$$Al(p,$$gamma$$)$$^{28}$$Si reaction. The measured response functions both with and without anti-coincidence were compared with the calculations by Monte Carlo simulations, in which the pulse-height degradation effect associated with a surface channel has been taken into account. Excellent agreements were obtained between the measured response functions and the simulations by introducing two fitting parameters for the BGO shields.