Refine your search:     
Report No.
 - 

Performance of a new signal processing system for a detector array

Kimura, Atsushi   ; Koizumi, Mitsuo ; Toh, Yosuke   ; Oshima, Masumi; Mizumoto, Motoharu; Goto, Jun*; Arai, Yasuo*; Sagara, Masahiro*; Iri, Shinichi*; Kobayashi, Hirotsugu*; Suzuki, Yasuhiro*

To measure neutron cross-section data of minor actinides, we construct a Ge-spectrometer utilizing multiple $$gamma$$-ray detection method. This spectrometer consists of 30 Ge crystals (40 detector outputs) and 128 BGO anti-Compton shields. Generally, a data acquisition system for such a big Ge spectrometer consists of many NIM modules; it requires large space and huge cost. To overcome these problems, we developed a new data acquisition system with digital signal processing techniques. We reported the conceptual design (and cost) in NSS 2004. In this NSS 2005, we report basic performance of this system (energy resolution, dead time and so on) and results of experiments with the Ge-spectrometer. We measured the energy resolution with a Eurisys clover detector, which is one component of the Ge-spectrometer. Obtained FWHMs (2.5 - 2.8 keV at 1.33 MeV) show good performances as normal MCA even with short shaping time and much higher channel density. The dead time of this system is only 3.2 micro sec per event.