Development of a neutron personal dose equivalent detector

Tsujimura, Norio   ; Yoshida, Tadayoshi ; Takada, Chie ; Momose, Takumaro  ; Nunomiya, Tomoya*; Aoyama, Kei*

The authors developed a new neutron measuring-instrument that was designed to measure neutron personal dose equivalent, $$H$$$$_{rm p}$$(10). This detector is composed a central thermal neutron detector, a core-polyethylene moderator, a front-hemispherical polyethylene moderator and rear-hemispherical boron-contained polyethylene absorber. Dimensions of each part were determined so as to provide energy- and angular-responses close to energy- and angular-dependencies of neutron $$H$$$$_{rm p}$$(10), based on the Monte Carlo calculations.