Performance of a data acquisition system for a large germanium detector array

Kimura, Atsushi   ; Koizumi, Mitsuo ; Toh, Yosuke   ; Goto, Jun*; Oshima, Masumi

We have evaluated performance of our new data acquisition system, which utilizes digital signal processing techniques and was developed for a 4$$pi$$ Ge spectrometer. Total energy resolution of this system was 2.5 keV in FWHM for 1.33-MeV $$gamma$$ rays and 16.4 keV in FWHM for 10.83-MeV prompt $$gamma$$ rays. The time resolution of timing signals, which are used for the measurements of coincident $$gamma$$ rays, was 120 ns in FWHM. Those values fulfilled our requirements for the DAQ system. A method of dead time correction with this system was also developed.