Refine your search:     
Report No.
 - 

A Half-millimetre spatial resolution fibre-coded linear position-sensitive scintillator detector with wavelength-shifting fibre read-out for neutron detection

Nakamura, Tatsuya  ; Schooneveld, E. M.*; Rhodes, N. J.*; Katagiri, Masaki; To, Kentaro  ; Sakasai, Kaoru  ; Soyama, Kazuhiko  

A fibre-coded linear scintillator detector with a half-millimetre spatial resolution was developed with wavelength-shifting fibre read-out. The head of the detector comprised a ZnS/$$^{6}$$LiF scintillator placed on the ribbon of WLS fibres. Both ends of the fibres were coded to reduce the number of read-out photomultipliers and associated electronic circuitry. The prototype detector in which the fibre ribbon was sandwiched between two ZnS/$$^{6}$$LiF scintillators exhibited a spatial resolution of 0.5 mm, a detection efficiency of 35% for 1-${AA}$ neutrons, and a $$gamma$$ sensitivity of less than $$10^{-7}$$. This type of detector would be useful where a submillimetre spatial resolution is required with large detector coverage such as in neutron reflectometers.

Accesses

:

- Accesses

InCites™

:

Percentile:66.94

Category:Instruments & Instrumentation

Altmetrics

:

[CLARIVATE ANALYTICS], [WEB OF SCIENCE], [HIGHLY CITED PAPER & CUP LOGO] and [HOT PAPER & FIRE LOGO] are trademarks of Clarivate Analytics, and/or its affiliated company or companies, and used herein by permission and/or license.