Refine your search:     
Report No.
 - 

Characteristics of convex hull data treatment dosimetry using multiple detectors

Hashimoto, Makoto ; Morev, M. N.*; Iimoto, Takeshi*; Kosako, Toshiso*

Convex hull data treatment dosimetry using multiple dosimeters is studied. It makes available to make reliable dosimetry with feedback of data about the radiation field derived from the relationship between multiple detector outputs. It holds that the combination of even simple detectors can evaluate the neutron personal dose equivalent or ambient dose equivalent in sufficient accuracy. The relation between the combination of detectors and the capability of dose evaluation is also discussed.