Measurement of microdosimetric spectra with a wall-less tissue-equivalent proportional counter for 290 MeV/u $$^{12}$$C beam

Tsuda, Shuichi  ; Sato, Tatsuhiko   ; Takahashi, Fumiaki  ; Satoh, Daiki   ; Endo, Akira   ; Sasaki, Shinichi*; Namito, Yoshihito*; Iwase, Hiroshi*; Ban, Shuichi*; Takada, Masashi*

The frequency distribution of the lineal energy of 290 MeV/u carbon beam was measured using a wall-less tissue equivalent proportional counter (wall-less TEPC) in a cylindrical volume with simulated diameter 0.72 $$mu$$m in verifying the accuracy of a dose calculation model. The measured lineal energy distribution as well as its dose-mean value agreed fairly well with the corresponding data from microdosimetric calculations using the PHITS code within the experimental uncertainty. It is found that a wall-less TEPC is needed to measure the precise energy deposition spectra of the delta rays produced secondarily by energetic heavy ion beams. The measured data also indicate that more than 11% of the energy escaped from the path of the trajectory of the carbon beam.