Yamaguchi, Mitsutaka; Kawachi, Naoki; Suzui, Nobuo; Fujimaki, Shu; Kamiya, Tomihiro; Odaka, Hirokazu*; Ishikawa, Shinnosuke*; Kokubun, Motohide*; Watanabe, Shin*; Takahashi, Tadayuki*; Shimada, Hirofumi*; Arakawa, Kazuo*; Suzuki, Yoshiyuki*; Torikai, Kota*; Yoshida, Yukari*; Nakano, Takashi*

We are constructing a three-dimensional imaging system for medical and biological applications. The system will allow simultaneous imaging at high spatial and energy resolutions across a wide energy range, from several tens of keV to a few MeV. In this work, one prototype head module have been developed for a multi-head Si/CdTe Compton camera system. The performance of the prototype was evaluated with a sealed Ba-133 radiation source. The experiments confirmed that the imaging results were consistent with actual source position. In addition to the resolution for the parallel directions to the detector surface, the position resolution was evaluated for the depth direction at a point in near region of the head-module. These position resolutions were well reproduced by Monte Carlo simulation results.