Development of an automatic prompt $$gamma$$-ray activation analysis system

Osawa, Takahito  

An automatic prompt $$gamma$$-ray activation analysis system was developed and installed at the Japan Research Reactor No.3 Modified (JRR-3M). This system is mainly composed of two personal computers, four programs, a six-axis vertical revolute joint robot, and a data acquisition device. The main control software, referred to as AutoPGA, was developed using LabVIEW 2011 and the hand-made program can control all functions of the analytical system. The core of the new system is an automatic sample exchanger and measurement system with several additional automatic control functions integrated into the system. Up to fourteen samples can be automatically measured by the system. Therefore, this system can not only dramatically increase the efficiency of routine measurements but also decrease the background level of $$gamma$$-rays for the analysis through the use of a sophisticated automatic control system.