Neutron-sensitive ZnS/$$^{10}$$B$$_{2}$$O$$_{3}$$ ceramic scintillator detector as an alternative to a $$^{3}$$He-gas-based detector for a plutonium canister assay system

Nakamura, Tatsuya  ; Ozu, Akira  ; To, Kentaro  ; Sakasai, Kaoru  ; Suzuki, Hiroyuki; Honda, Katsunori; Birumachi, Atsushi; Ebine, Masumi ; Yamagishi, Hideshi*; Takase, Misao; Haruyama, Mitsuo; Kureta, Masatoshi   ; Soyama, Kazuhiko  ; Nakamura, Hironobu ; Seya, Michio

A neutron-sensitive ZnS/$$^{10}$$B$$_{2}$$O$$_{3}$$ ceramic scintillator detector was developed as an alternative to a $$^{3}$$He-gas-based detector for use in a plutonium canister assay system. The detector has a modular structure, with a flat ZnS/$$^{10}$$B$$_{2}$$O$$_{3}$$ceramic scintillator strip that is installed diagonally inside a light-reflecting aluminium case with a square cross section. The prototype detectors, which have a neutron-sensitive area of 30 mm $$times$$ 250 mm, exhibited a sensitivity of 21.7-23.4 $$pm$$ 0.1 cps$$/$$nv for thermal neutrons, a $$^{137}$$Cs $$gamma$$-ray sensitivity of 1.1-1.9 $$pm $$0.2 $$times$$ 10$$^{-7}$$ and a count variation of less than 6% over the detector length. A trial experiment revealed a temperature coefficient of less than -0.24$$pm$$ 0.05% / $$^{circ}$$C over the temperature range of 20-50$$^{circ}$$C.