Refine your search�ソスF     
Report No.
 - 

Verification of alternative dew point hygrometer for CV-LRT in MONJU; Short and long-term verification for capacitance type dew point hygrometer

Hatori, Masakazu; Ichikawa, Shoichi; Chiba, Yusuke; Kobayashi, Takanori; Uekura, Ryoichi; Hashiri, Nobuo*; Inuzuka, Taisuke*; Kitano, Hiroshi*; Abe, Hisashi*; Ono, Fumiyasu

In order to reduce the influence on a plant schedule of Monju by the maintenance of dew point hygrometers, JAEA Agency examined a capacitance type dew point hygrometer as an alternative dew point hygrometer for a lithium-chloride type dew point hygrometer which had been used at the containment vessel leak rate test in Monju. As a result of comparing a capacitance type dew point hygrometer with a lithium-chloride type dew point hygrometer at the CV-LRT (Atmosphere: nitrogen, testing time/24 hours), there was no significant difference between a capacitance type dew point hygrometer and a lithium-chloride type dew point hygrometer. As a result of comparing a capacitance dew point hygrometer with a high-precision-mirror-surface type dew point hygrometer for long term verification (Atmosphere: air, testing time: 24 months), it was confirmed that a capacitance type dew point hygrometer satisfied the instrument specification required by the CV-LRT. There are no other cases where atmosphere measurement has been carried out for a long time by existing measuring instruments.