A Cubic CeBr$$_{3}$$ gamma-ray spectrometer suitable for the decommissioning of the Fukushima Daiichi Nuclear Power Station

Kaburagi, Masaaki   ; Shimazoe, Kenji*; Otaka, Yutaka*; Uenomachi, Mizuki*; Kamada, Kei*; Kim, K. J.*; Yoshino, Masao*; Shoji, Yasuhiro*; Yoshikawa, Akira*; Takahashi, Hiroyuki*; Torii, Tatsuo  

Our work focused on the passive gamma-ray analysis (PGA) of the nuclear fuel debris based on measuring gamma rays with an energy greater than 1 MeV for the decommissioning of the Fukushima Daiichi Nuclear Power Station (FDNPS). The PGA requires gamma-ray spectrometers to be used under the high dose rates in the FDNPS, then we fabricated a small cubic CeBr$$_{3}$$ spectrometer with dimensions of 5 mm $$times$$ 5 mm $$times$$ 5 mm, coupled to a Hamamatsu R7600U-200 photomultiplier tube (PMT). The performance at dose rates of 4.4 to 750 mSv/h in a $$^{60}$$Co field was investigated. The energy resolution (FWHM) at 1333 keV ranged from 3.79% to 4.01%, with a standard deviation of 6.9%, which met the narrow gamma decay spectral lines between $$^{154}$$Eu (1274 keV) and $$^{60}$$Co (1333 keV). However, the spectra shifted to a higher energy level as the dose rate increase, there was a 51% increase at the dose rates of 4.4 to 750 mSv/h, which was caused by the PMT gain increase.