High-intensity beam profile measurement using a gas sheet monitor by beam induced fluorescence detection

Yamada, Ippei; Wada, Motoi*; Moriya, Katsuhiro  ; Kamiya, Junichiro   ; Saha, P. K.   ; Kinsho, Michikazu  

A transverse beam profile monitor that visualizes a two-dimensional beam-induced fluorescent image was developed. The monitor employs a sheet-shaped gas flow formed by a technique of rarefied gas dynamics. A simplified analysis method was developed to reconstruct the beam intensity profile from the obtained image. The developed profile monitor and the analysis method were applied to measure the J-PARC 3 MeV H$$^-$$ beam profile. The root mean square values of the profiles were consistent with the ones obtained by a wire-scanning-type beam profile monitor. The beam loss due to the gas sheet injection was measured as a beam-current reduction. The amount of the beam current decreased in proportion to the gas sheet flux and the reduction ranged from 0.004 to 2.5%. The assembled system was capable of reconstructing a beam profile from a single shot beam pulse (1.7$$times$$10$$^{13}$$ protons in 50 $$mu$$s).