Estimation of reaction constants of actinide ions with O$$_{2}$$ in reaction cell in triple quadrupole inductively coupled plasma mass spectrometry

Kazama, Hiroyuki  ; Konashi, Kenji*; Suzuki, Tatsuya*; Sekio, Yoshihiro  ; Maeda, Koji  ; Koyama, Shinichi  ; Abe, Chikage*; Nagai, Yasuyoshi*

Ultratrace analysis technique is important to understand fuel debris in nuclear reactor core after sever accident. Ion-molecule reactions of actinides ($$^{237}$$Np, $$^{241}$$Am and $$^{244}$$Cm) in a reaction cell were measured by inductively coupled plasma mass spectrometry (ICP-MS/MS). These nuclides are included in fuel debris. A gas-phase ion-molecule reaction model has been developed to simulate the gas phase reactions in the reaction cell. The results of model simulation are good agreement with the flow rates dependences of experimental data accurately.