Cavity ionization chamber responses in the JAEA and the NMIJ high-energy photon reference fields for radiation protection

放射線防護に用いるJAEAとNMIJの高エネルギー光子標準場に対する空洞電離箱の応答評価

石井 隼也*; 清水 森人*; 加藤 昌弘*; 黒澤 忠弘*; 辻 智也   ; 吉富 寛   ; 谷村 嘉彦   ; 渡部 浩司*

Ishii, Junya*; Shimizu, Morihito*; Kato, Masahiro*; Kurosawa, Tadahiro*; Tsuji, Tomoya; Yoshitomi, Hiroshi; Tanimura, Yoshihiko; Watabe, Hiroshi*

The dosimeter response should be calibrated in a reference field near the user's radiation environment. Environments around nuclear reactors and radiation therapy facilities have high-energy photons with energies exceeding that of $$^{60}$$Co gamma rays, and controlling exposure to these photons is important. The Japan Atomic Energy Agency (JAEA) and National Metrology Institute of Japan (NMIJ) have high-energy reference fields with energies above several MeV for different types of accelerators. Their reference fields have different fluence-energy distributions. In this study, the energy dependencies of the two-cavity ionization chambers, which are often used by secondary laboratories, were experimentally and computationally evaluated for each high-energy field. These results agreed well with relative expanded uncertainties ($$k$$ = 2), and their capabilities for air kerma measurements in each high-energy reference field were confirmed. Therefore, the capabilities of the air-kerma measurements can be verified in the two high-energy reference fields.

Access

:

- Accesses

InCites™

:

パーセンタイル:0.00

分野:Environmental Sciences

Altmetrics

:

[CLARIVATE ANALYTICS], [WEB OF SCIENCE], [HIGHLY CITED PAPER & CUP LOGO] and [HOT PAPER & FIRE LOGO] are trademarks of Clarivate Analytics, and/or its affiliated company or companies, and used herein by permission and/or license.