Cavity ionization chamber responses in the JAEA and the NMIJ high-energy photon reference fields for radiation protection
放射線防護に用いるJAEAとNMIJの高エネルギー光子標準場に対する空洞電離箱の応答評価
石井 隼也*; 清水 森人*; 加藤 昌弘*; 黒澤 忠弘*; 辻 智也
; 吉富 寛
; 谷村 嘉彦
; 渡部 浩司*
Ishii, Junya*; Shimizu, Morihito*; Kato, Masahiro*; Kurosawa, Tadahiro*; Tsuji, Tomoya; Yoshitomi, Hiroshi; Tanimura, Yoshihiko; Watabe, Hiroshi*
The dosimeter response should be calibrated in a reference field near the user's radiation environment. Environments around nuclear reactors and radiation therapy facilities have high-energy photons with energies exceeding that of
Co gamma rays, and controlling exposure to these photons is important. The Japan Atomic Energy Agency (JAEA) and National Metrology Institute of Japan (NMIJ) have high-energy reference fields with energies above several MeV for different types of accelerators. Their reference fields have different fluence-energy distributions. In this study, the energy dependencies of the two-cavity ionization chambers, which are often used by secondary laboratories, were experimentally and computationally evaluated for each high-energy field. These results agreed well with relative expanded uncertainties (
= 2), and their capabilities for air kerma measurements in each high-energy reference field were confirmed. Therefore, the capabilities of the air-kerma measurements can be verified in the two high-energy reference fields.