Taguchi, Shigeo; Miyauchi, Hironari*; Horigome, Kazushi; Yamamoto, Masahiko; Kuno, Takehiko
Bunseki Kagaku, 67(11), p.681 - 686, 2018/11
In thermal ionization mass spectrometry, de-gassing is one of the important treatments to release impurities of filaments and to minimize the influence of background. In this work, the effect of the surface change in the tungsten filament induced by the conductively heating treatment on uranium isotopic (U/U) measurement has been investigated. It was found that the conductively heating treatment of the filament has the effect of smoothing the surface of the filament and also has the effect of improving the deposition of the sample on the filament surface. As a result of either these effects, the precision of uranium isotopic (U/U) measurement was improved.
Masaki, Yuji*; Miyauchi, Hironari*; Taguchi, Shigeo; Horigome, Kazushi; Ishibashi, Atsushi; Yamamoto, Masahiko; Masui, Kenji; Nishida, Naoki; Kuno, Takehiko; Surugaya, Naoki
no journal, ,
Reprocess the technology development center has longtime experience for the handling of the radiation and perform support aiming at the acquisition of the spot analysis technique about the radioactive nuclear material handling in response to a demand out of the mechanism.