高橋 聖; 関口 真人; 宮内 英明; 橘 晴夫; 吉澤 道夫; 加藤 徹*; 山口 明仁*
Journal of Nuclear Science and Technology, 45(Suppl.5), p.225 - 228, 2008/06
被引用回数:5 パーセンタイル:40.16(Nuclear Science & Technology)In this study, the H(10) and H(0.07) detectable electronic pocket dosimeter(EPD-101) manufactured by ALOKA CO., LTD. was examined for (1) energy dependences for and -rays and (2) applicability for actually personal dosimetry in the mixed radiation fields. The energy responses of EPD-101 were obtained by standard irradiation in the Facility of Radiation Standards in the Japan Atomic Energy Agency (JAEA). The responses of EPD-101 were shown to be within 20% in the irradiated energy range for the both radiations except for Pm emitting low-energy -rays. In order to examine the applicability of EPD-101 in actual radiation workplaces, measured values of EPD-101 were compared with those of two other types of dosimeters (glass dosimeters and thermoluminescence dosimeters) in hot cells of the Reactor Fuel Examination Facility in JAEA. As a result, EPD-101 showed lower dose for -ray and higher dose for -ray than the other two dosimeters. The differences should be attributed to high-energy -emitting sources such as Ru-Rh in the facility.
高橋 聖; 関口 真人; 橘 晴夫; 星 慎太郎; 吉澤 道夫; 加藤 徹*; 山口 明仁*
no journal, ,
従来、原子力科学研究所では線混合場の作業の線量管理には熱ルミネセンス線量計(TLD)を補助線量計として多く使用してきた。今回、線同時測定が可能なアロカ社製70m線量当量対応形電子ポケット線量計(EPD)について、線に対するエネルギー特性試験及び実作業環境場における70m線量当量のTLD(UD-808型)との比較測定を行った。EPDの線エネルギー特性試験は原科研放射線標準施設の線標準照射場において行った。実作業環境場でのEPDとTLDの比較は、使用済燃料取扱施設でEPD, TLD及び原科研において被ばく管理上の基本線量計として使用しているガラス線量計を同時に胸部に装着し、各線量計による70m線量当量を測定して行った。EPDの線のエネルギー特性は、残留最大エネルギー0.5MeV2.2MeVで30%の許容範囲内に収まっていることが確認できた。また、実作業環境場での70m線量当量の測定結果について、EPDの方がTLDよりも基本線量計であるガラス線量計に近い指示値を示した。EPDは、線量の表示や内蔵メモリーによる時系列データの取込み等の機能面に優れており、また取扱が容易であることから、線混合場の作業における線量管理について有用であると考えられる。