Verification of alternative dew point hygrometer for CV-LRT in MONJU; Short- and long-term verification for capacitance-type dew point hygrometer (Translated document)

Ichikawa, Shoichi; Chiba, Yusuke; Ono, Fumiyasu; Hatori, Masakazu; Kobayashi, Takanori; Uekura, Ryoichi; Hashiri, Nobuo*; Inuzuka, Taisuke*; Kitano, Hiroshi*; Abe, Hisashi*

JAEA-Research 2017-001, 40 Pages, 2017/03

JAEA-Research-2017-001.pdf:5.19MB

In order to reduce the influence on a plant schedule of the MONJU by the maintenance of dew point hygrometers, The JAEA examined a capacitance type dew point hygrometer as an alternative dew point hygrometer for a lithium-chloride type dew point hygrometer which had been used at the CV-LRT in the MONJU. As a result of comparing a capacitance type dew point hygrometer with a lithium-chloride type dew point hygrometer at the CV-LRT (Atmosphere: nitrogen, Testing time: 24 hours), there weren't significant difference between a capacitance type dew point hygrometer and a lithium-chloride type dew point hygrometer. As a result of comparing a capacitance dew point hygrometer with a high-mirror-surface type dew point hygrometer for long term verification (Atmosphere: air, Testing time: 24 months), the JAEA confirmed that a capacitance type dew point hygrometer satisfied the instrument specification ($$pm$$2.04$$^{circ}$$C) required by the JEAC4203-2008.

JAEA Reports

Verification of alternative dew point hygrometer for CV-LRT in Monju

Ichikawa, Shoichi; Chiba, Yusuke; Ono, Fumiyasu; Hatori, Masakazu; Kobayashi, Takanori; Uekura, Ryoichi; Hashiri, Nobuo*; Inuzuka, Taisuke*; Kitano, Hiroshi*; Abe, Hisashi*

JAEA-Research 2016-021, 32 Pages, 2017/02

JAEA-Research-2016-021.pdf:5.0MB

In order to reduce the influence on a plant schedule of the MONJU by the maintenance of dew point hygrometers, The JAEA examined a capacitance type dew point hygrometer as an alternative dew point hygrometer for a lithium-chloride type dew point hygrometer which had been used at the CV-LRT in the MONJU. As a result of comparing a capacitance type dew point hygrometer with a lithium-chloride type dew point hygrometer at the CV-LRT (Atmosphere: nitrogen, Testing time: 24 hours), there weren't significant difference between a capacitance type dew point hygrometer and a lithium-chloride type dew point hygrometer. As a result of comparing a capacitance dew point hygrometer with a high-mirror-surface type dew point hygrometer for long term verification (Atmosphere: air, Testing time: 24 months), the JAEA confirmed that a capacitance type dew point hygrometer satisfied the instrument specification ($$pm$$2.04$$^{circ}$$C) required by the JEAC4203-2008.

Oral presentation

Verification test of alternative dew point detector for CV-LRT in Monju

Chiba, Yusuke; Ichikawa, Shoichi; Ono, Shimpei; Hatori, Masakazu; Kobayashi, Takanori; Uekura, Ryoichi; Hashiri, Nobuo*; Inuzuka, Taisuke*; Abe, Hisashi*; Kitano, Hiroshi*

no journal, , 

The lithium chloride-type dew point detector was used for reactor containment vessel the entire leak rate test (CV-LRT) in fast-breeder reactor Monju (Monju) which needs the maintenance every three months. But when the plant process is considered, it's desirable for the maintenance periods beyond 12 months. Thereupon the capacitance-type dew point detector made in VAISALA Corporation was nominated as the alternative lithium chloride-type dew point detector in Monju. The lithium chloride-type dew point detector, when it's adopted, for, to have to be satisfied a request of the regulations, it's necessary to estimate the performance of the lithium chloride-type dew point detector. In the condition of the CV-LRT, the dew point data were measured by the capacitance-type dew point detector and the lithium chloride-type dew point detector for 24 hours. These data were compared and the performance of the capacitance-type dew point detector was estimated. Furthermore, the dew point data of Monju under the atmosphere ware continuously measured by the capacitance-type dew point detector and the mirror surface-type dew point detector for 2 years. These data were compared and the performance of the capacitance-type dew point detector was estimated. As a result, the capacitance-type dew point detector was confirmed that had the instrument precision that JEAC4203-2008 required.

Oral presentation

Verification of alternative dew point hygrometer for CV-LRT in MONJU; Short and long-term verification for capacitance type dew point hygrometer

Hatori, Masakazu; Ichikawa, Shoichi; Chiba, Yusuke; Kobayashi, Takanori; Uekura, Ryoichi; Hashiri, Nobuo*; Inuzuka, Taisuke*; Kitano, Hiroshi*; Abe, Hisashi*; Ono, Fumiyasu

no journal, , 

In order to reduce the influence on a plant schedule of Monju by the maintenance of dew point hygrometers, JAEA Agency examined a capacitance type dew point hygrometer as an alternative dew point hygrometer for a lithium-chloride type dew point hygrometer which had been used at the containment vessel leak rate test in Monju. As a result of comparing a capacitance type dew point hygrometer with a lithium-chloride type dew point hygrometer at the CV-LRT (Atmosphere: nitrogen, testing time/24 hours), there was no significant difference between a capacitance type dew point hygrometer and a lithium-chloride type dew point hygrometer. As a result of comparing a capacitance dew point hygrometer with a high-precision-mirror-surface type dew point hygrometer for long term verification (Atmosphere: air, testing time: 24 months), it was confirmed that a capacitance type dew point hygrometer satisfied the instrument specification required by the CV-LRT. There are no other cases where atmosphere measurement has been carried out for a long time by existing measuring instruments.