Mehta, K.*; Kojima, Takuji; Sunaga, Hiromi
Radiation Physics and Chemistry, 68(6), p.959 - 962, 2003/12
Times Cited Count:3 Percentile:25.79(Chemistry, Physical)Applicability of the existing dosimetry systems to high-power Bremsstrahlung irradiation was investigated through a dose intercomparison study, where several dosimeters were irradiated in the dose range 4-12 kGy in identical polyethylene phantoms in a Bremsstrahlung beam obtained from a 5-MeV electron accelerator. Included in the study were alanine dosimeters molded by three different binders, three types of liquid dosimeters:ceric-cerous, dichromate and ethanol-chlorobenzen (ECB), and glutamine powder. The dosimeter responses for Bremsstrahlung radiation were analyzed at the issuing laboratories, and the dose values determined using calibration based on cobalt-60 -ray irradiation. Dose values for all the three dose levels for all dosimetry systems were in good agreement of better than 3%. The results of the study demonstrate that these existing dosimetry systems have a potential for application to high-power Bremsstrahlung irradiation.