Nakamura, Tatsuya; Katagiri, Masaki; Chen, Y. E.*; Ukibe, Masahiro*; Okubo, Masataka*
Nuclear Instruments and Methods in Physics Research A, 559(2), p.766 - 768, 2006/04
Times Cited Count:13 Percentile:64.62(Instruments & Instrumentation)We have been developing a position-sensitive neutron detector, which exhibits a high spatial resolution of the order of 10 microns and high detection efficiency. The analysis on the pulse-heights or on the time-delay between the STJs gives incident positions of the neutrons. In this paper, the response of the STJs/LBO neutron detector measured using a cold neutron beam is presented. The two STJs have the same size of 50 50
m
and their distance is 1.3 mm. The count intensity in the correlation map varied depending on the incident position of the neutron beam which gives an evidence for the position-sensitivity of the detector. These experimental results were qualitatively understood by a simple solid-angle model.