東條 寛; 山田 一博*; 安原 亮*; 谷塚 英一; 舟場 久芳*; 波多江 仰紀; 林 浩*; 伊丹 潔
Review of Scientific Instruments, 85(11), p.11D865_1 - 11D865_3, 2014/11
被引用回数:2 パーセンタイル:10.84(Instruments & Instrumentation)In incoherent Thomson scattering diagnostics, stray light of the incident laser, which is frequently generated at the vacuum windows, also causes wrong evaluations of the signal intensities. Because the laser goes through the vacuum windows many times in a double-pass (or multi-pass) scattering system, complicate temporal structures can be seen in the output signals. In this paper, Singular Value Decomposition (SVD) is applied to the output signals in a Thomson scattering diagnostic on Large Helical Device (LHD). The analyzed data were all measured signals for each one spectral channel over one plasma shot. Some noises were filtered and Te was evaluated within good accuracies especially in the second pass (forward scattering). We compared the errors in Te (0.8 - 1.3 keV) from the noise-filtered and from non-filtered data. Use of the filtered data enables suppressing the error in Te measured from the second pass by a factor of 0.5 at the most.
篠原 孝司; 石井 啓一*; 落合 謙太郎; 馬場 護*; 助川 篤彦; 笹尾 眞實子*; 北島 純男*
Review of Scientific Instruments, 85(11), p.11E823_1 - 11E823_4, 2014/11
被引用回数:0 パーセンタイル:0.00(Instruments & Instrumentation)A neutron flux array is one of the important plasma diagnostics to obtain the information on fast ion population. However, the limited flux owing to its collimator results in the small number of pulse counts and the large statistical error. The method to increase the pulse counts is required. Here, we propose a multi-stage neutron detection system, in which several sets of a scintillator and photo-multiplier tube (PMT) are placed in a line-of-sight. In order to evaluate the performance of the multi-stage detection system, we have carried out experiments on a two-stage detection system using a neutron beam at FNS (Fusion Neutronics Source) of JAEA (Japan Atomic Energy Agency). The results have shown the concept of the multi-stage detection system works as expected. In the best setup, the test two-stage system reached about 1.7 (1.8) times the efficiency of a single scintillator and PMT system for 2.54 (14) MeV neutrons.
内田 雄大*; 高田 英治*; 藤崎 明広*; 磯部 光孝*; 小川 国大*; 篠原 孝司; 富田 英生*; 河原林 順*; 井口 哲夫*
Review of Scientific Instruments, 85(11), p.11E118_1 - 11E118_4, 2014/11
被引用回数:9 パーセンタイル:38.80(Instruments & Instrumentation)Neutron and -ray discrimination with a digital signal processing system has been used to measure the neutron emission profile in magnetic confinement fusion devices. However, a sampling rate must be set low to extend the measurement time because the memory storage is limited. Time jitter decreases a discrimination quality due to a low sampling rate. As described in this paper, a new charge comparison method was developed. Furthermore, automatic neutron- discrimination method was examined using a probabilistic approach. Analysis results were investigated using the figure of merit. Results show that the discrimination quality was improved. Automatic discrimination was applied using the EM algorithm and k-means algorithm.