Ito, Hirokuni*; Hatakeyama, Mutsuo*; Tachibana, Mitsuo; Yanagihara, Satoshi
JAERI-Tech 2003-012, 34 Pages, 2003/03
The MISE was developed to evaluate low-level radiological contaminations of inner surfaces of piping. The MISE consists of a cylindrically-formed double layered type detector and a piping crawling robot, which were designed and manufactured separately. In measurements of the contaminations, an outer cylindrical detector close to the surface of piping measures -rays and -rays and an inner cylindrical detector set after a shielding plate for shield of -rays measures -rays. The -ray counting rates are derived by subtracting -ray counts measured by the inner detector from - and -ray counts measured by the outer detector. The robot transports the double layered type detector with observing inner surfaces of piping. The detection limit for the contamination of Co was found to be about 0.17Bq/cm with measurement time of 30 seconds. It is expected that 0.2Bq/cm corresponding to clearance level of Co (0.4Bq/g) can be evaluated with measurement time of 2 seconds, which is equal to measurement speed of 54m/h.