Li-glass detector to gamma rays by a coincidence method伊藤 史哲*; Lee, J.; 弘中 浩太; 高橋 時音; 鈴木 敏*; 持丸 貴則*; 堀 順一*; 寺田 和司*; 小泉 光生
Nuclear Instruments and Methods in Physics Research A, 1064, p.169465_1 - 169465_9, 2024/07
被引用回数:0 パーセンタイル:0.00(Instruments & Instrumentation)The response of a gamma-ray spectrometer is generally determined by analyzing full-energy peaks. However, full-energy peaks cannot be measured easily in the case of scintillation detectors that consist of light elements, such as glass scintillators. Only a strong Compton plateau appears in the spectrum of such detectors. Therefore, Compton edgers were used to evaluate the response of these detectors. The response of a low-resolution
Li-glass detector to gamma rays was measured for the first time by a coincidence method with a high-resolution LaBr
:Ce detector using cascade gamma rays (2.75 and 1.37 MeV) from a
Na source. Coincidence gates were applied at the peaks of the spectrum of the LaBr
:Ce detector at the 0.51 MeV annihilation peak, and the sum peaks of a gamma ray and a backscattered gamma ray. By analyzing the gated spectra of the
Li-glass detector, the energy-dependent detector response (i.e., the output strength and its dispersion) was determined.