Nakamura, Tatsuya; To, Kentaro; Ebine, Masumi; Birumachi, Atsushi; Sakasai, Kaoru
Proceedings of 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC 2019), Vol.1, p.735 - 736, 2020/08
Times Cited Count:0 Percentile:0.00(Nuclear Science & Technology)A large area, position-sensitive scintillation neutron detector was developed for upgrading the SENJU, time-of-flight Laue single crystal neutron diffractometer, in J-PARC MLF. The detector has a neutron-sensitive area of 512
512 mm with a pixel size of 4
4 mm. The detector was developed for upgrading of the SENJU instrument. The large area detector is to be installed below the vacuum tank to enlarge a covering solid angle. A
Li:ZnS (Ag) scintillator and wavelength-shifting fiber technologies are employed. Each fiber channel is read out individually with photon counting mode. The electronics boards are implemented at the backside of the detector, enabling the detector depth as short as 20 cm. The detector exhibited a detection efficiency of 45% for thermal neutron. No degradation in fiber position and in neutron sensitivity has been observed over one year after production. In this paper, detector design and detector performances are presented.
Nakamura, Tatsuya; To, Kentaro; Tsutsui, Noriaki; Ebine, Masumi; Birumachi, Atsushi
no journal, ,
A position-sensitive scintillation detector module for a new protein neutron diffractometer was developed by using a scintillator / wavelength shifting (WLS) fiber technology. The detector module has a spatial resolution of 2.5 mm with a neutron-sensitive area of 320
320 mm
. The WLS fibers are arranged in a regular pitch of 2.5 mm in x and y direction and those arrays are placed diagonally. The light reflecting grid is inserted in between the fibers in order for optical isolation and for mechanical support of the fiber. The detector implemented flat
Li/ZnS screens up and downstream of the WLF fiber arrays. The detector exhibited a detection efficiency of 30-50% for thermal neutron (depending on
-ray sensitivities) and a count uniformity of ~13%. In this paper detailed detector design and experimental results obtained using a pulsed neutron beam are presented.