Ishikawa, Masayori*; Ono, Koji*; Matsumura, Akira*; Yamamoto, Tetsuya*; Hiratsuka, Junichi*; Miyatake, Shinichi*; Kato, Itsuro*; Sakurai, Yoshinori*; Kobayashi, Toru*; Kumada, Hiroaki; et al.
Proceedings of 12th International Congress on Neutron Capture Therapy (ICNCT-12), p.397 - 400, 2006/10
An ultraminiature thermal neutron monitor which was named SOF detector (Scintillator with Optical Fiber detector) had been developed for BNCT treatment. We had been experienced 15 clinical trials using SOF detector until the end of 2005, some measurements got good results, and some got unacceptable results. One reason of the unacceptable results was due to dislocation of the detector during treatment. This is because it is difficult to fix the SOF detector on patient's skin without strong sticker. To overcome this problem, a loop-type SOF probe was developed. By using the loop-type SOF detector, fixing on the patient's skin was much easier.