Study of chemical etching conditions for alpha-particle detection and visualization using solid state nuclear track detectors

Yamada, Ryohei; Odagiri, Taiki*; Iwaoka, Kazuki*; Hosoda, Masahiro*; Tokonami, Shinji*

Radiation Environment and Medicine, 8(1), p.21 - 25, 2019/02

We evaluate radon/thoron and its progeny concentration using passive-type monitors using CR-39 plates. After exposure, it is necessary to do chemical etching for CR-39 plates. In the present study, we considered shortening of chemical etching time for CR-39 and enlargement of the track diameter (i.e. etch pit diameter) aiming for introduction of automatic counting system in the future. Optimum conditions were determined by changing solution concentration, solution temperature and etching time. As a result, the optimized conditions (concentration, temperature and etching time) were determined to be 8 M NaOH solution, 75 degrees Celsius and 10 hours. This result of etching time showed that the chemical etching was completed in less than half of conventional etching time. Furthermore, it was suggested that shorter etching time would be possible if we do not consider the enlargement of conventional track diameter.