Enhancement in dose sensitivity of polymer gel dosimeters composed of radiation-crosslinked gel matrix and less toxic monomers

Hiroki, Akihiro; Yamashita, Shinichi*; Taguchi, Mitsumasa

Polymer gel dosimeters based on radiation-crosslinked hydroxypropyl cellulose gel were prepared, which comprised 2-hydroxyethyl methacrylate (HEMA) and polyethylene glycol dimethacrylate (9G) as less toxic monomers and tetrakis (hydroxymethyl) phosphoniumu chloride (THPC) as an antioxidant. The dosimeters exposed to $$^{60}$$Co $$gamma$$-rays became cloudy at only 1 Gy. The irradiated dosimeters were optically analyzed by using a UV-vis spectrophotometer to evaluate dose response. Absorbance of the dosimeters linearly increased in the dose range from 0 to 10 Gy. The dose sensitivity increased with increasing 9G concentration. The dosimeter with 2 wt% HEMA, 3 wt% 9G, and 0.16 wt% THPC exhibited about 0.04 Abs Gy$$^{-1}$$ of the dose sensitivity. The dose sensitivity of the dosimeters was also enhanced by increment in THPC, reached about 0.06 Abs Gy$$^{-1}$$ at 0.40 wt% THPC.