Dose measurement precision of an RPLD-based eye lens dosemeter applicable to the medical sector

眼の水晶体モニタリングに用いるガラス線量計の校正定数の決定方法

古渡 意彦*; 永元 啓介*; 中上 晃一*; 吉富 寛   ; 盛武 敬*; 欅田 尚樹*

Kowatari, Munehiko*; Nagamoto, Keisuke*; Nakagami, Koichi*; Yoshitomi, Hiroshi; Moritake, Takashi*; Kunugita, Naoki*

We demonstrate a practical calibration method and its applicability for a commercially available radiophotoluminescence dosemeter (RPLD, GD-352M) to eye lens dose monitoring, by performing the calibration in accord with the ISO recommendations. The calibration was then verified through a series of experiments. For verification of the derived calibration factor (1.21 $$pm$$ 0.04, $$k$$ = 1) of the RPLD, we performed standard irradiations in the X-ray reference fields and the mockup measurements in the actual radiation fields in a hospital. The commercially available dosemeters, DOSIRIS was also put on the ISO cylinder phantom and the RANDO phantom together with the GD-352M in the verification experiments. The personal dose equivalents $$H_{rm p}$$(3) obtained from the GD-352M and those obtained from the DOSIRIS were in good agreement with each other. Our results demonstrate the proper calibration of a commercially available RPLD that is applicable to the additional monitoring of the lens of the eyes for medical personnel.