検索対象:     
報告書番号:
※ 半角英数字
 年 ~ 
 年

Calculating off-axis efficiency of coaxial HPGe detectors by Monte Carlo simulation

モンテカルロシミュレーションによるの同軸HPGe検出器の軸外効率の計算

Omer, M.  ; 静間 俊行*; 羽島 良一*; 小泉 光生 

Omer, M.; Shizuma, Toshiyuki*; Hajima, Ryoichi*; Koizumi, Mitsuo

In beam geometries where a directed $$gamma$$-ray beam hits the surface of a coaxial high purity germanium detector (HPGe), the detector efficiency is sensitive to the position where $$gamma$$-rays initially hit the detector surface because the structure of the detector is nonuniform. This may cause inaccuracy of the detector efficiency when measured using standard sources that are point-like sources emitting $$gamma$$-rays isotropically. Obtaining a precise estimation of the full energy peak efficiency of the coaxial HPGe detector in the beam geometry for on-axis and off-axis measurements requires a Monte Carlo simulation. We performed a Monte Carlo simulation that calculates the detector efficiency in the beam geometry. The effects of the off-axis distance and $$gamma$$-ray beam size on the efficiency are quantitatively analyzed. We found that the intrinsic efficiency in the beam geometry is maximized when the beam hits the detector at specific off-axis distances. Our Monte Carlo calculations have been supported by a nuclear resonance fluorescence experiment using laser Compton scattering $$gamma$$-ray beam.

Access

:

- Accesses

InCites™

:

パーセンタイル:53.91

分野:Chemistry, Physical

Altmetrics

:

[CLARIVATE ANALYTICS], [WEB OF SCIENCE], [HIGHLY CITED PAPER & CUP LOGO] and [HOT PAPER & FIRE LOGO] are trademarks of Clarivate Analytics, and/or its affiliated company or companies, and used herein by permission and/or license.