Nuclear resonant scattering measurements on $$^{57}$$Fe by multichannel scaling with a 64 pixel silicon avalanche photodiode linear-array detector

64chピクセルAPD検出器を用いた核共鳴散乱測定

岸本 俊二*; 三井 隆也; 春木 理恵*; 依田 芳卓*; 谷口 敬*; 島崎 昇一*; 池野 正弘*; 斉藤 正俊*; 田中 真伸*

Kishimoto, Shunji*; Mitsui, Takaya; Haruki, Rie*; Yoda, Yoshitaka*; Taniguchi, Takashi*; Shimazaki, Shoichi*; Ikeno, Masahiro*; Saito, Masatoshi*; Tanaka, Manobu*

We developed a silicon avalanche photodiode (Si-APD) linear-array detector for use in nuclear resonant scattering experiments using synchrotron X-rays. The Si-APD linear array consists of 64 pixels (pixel size: 0,01$$times$$0.02 cm$$^{2}$$) with a pixel pitch of 0.015 cm and depletion depth of 0.001 cm. An ultrafast frontend circuit allows the X-ray detector to obtain a high output. High-performance integrated circuits achieve multichannel scaling over 1024 continuous time bins with a 1 ns resolution for each pixel without dead time. The multichannel scaling method enabled us to record a time spectrum of the 14.4 keV nuclear radiation at each pixel with a time resolution of 1.4 ns (FWHM). This method was successfully applied to nuclear forward scattering and nuclear small-angle scattering on $$^{57}$$Fe.

Access

:

- Accesses

InCites™

:

パーセンタイル:38.80

分野:Instruments & Instrumentation

Altmetrics

:

[CLARIVATE ANALYTICS], [WEB OF SCIENCE], [HIGHLY CITED PAPER & CUP LOGO] and [HOT PAPER & FIRE LOGO] are trademarks of Clarivate Analytics, and/or its affiliated company or companies, and used herein by permission and/or license.