検索対象:     
報告書番号:
※ 半角英数字
 年 ~ 
 年

Development of a neutron beam monitor with a thin plastic scintillator for nuclear data measurement using spallation neutron source

中野 秀仁*; 片渕 竜也*; Rovira Leveroni, G.   ; 児玉 有*; 寺田 和司*; 木村 敦   ; 中村 詔司  ; 遠藤 駿典   

Nakano, Hideto*; Katabuchi, Tatsuya*; Rovira Leveroni, G.; Kodama, Yu*; Terada, Kazushi*; Kimura, Atsushi; Nakamura, Shoji; Endo, Shunsuke

A neutron monitoring detection system was developed for neutron capture cross section measurement using a spallation neutron source. A combination of a plastic scintillator and a thin $$^{6}$$LiF foil was adopted for the detector. The detector system was tested to study the feasibility of the system. Neutron irradiation experiments were conducted with the Accurate Neutron-Nucleus Reaction Measurement Instrument in the Materials and Life Science facility of the Japan Proton Accelerator Research Complex. A neutron time-of-flight spectrum was successfully measured without significant count loss or detector paralysis. The statistical uncertainty reached 0.7% at neutron energies around 6 meV.

Access

:

- Accesses

InCites™

:

パーセンタイル:31.61

分野:Nuclear Science & Technology

Altmetrics

:

[CLARIVATE ANALYTICS], [WEB OF SCIENCE], [HIGHLY CITED PAPER & CUP LOGO] and [HOT PAPER & FIRE LOGO] are trademarks of Clarivate Analytics, and/or its affiliated company or companies, and used herein by permission and/or license.