A Sensitive method for Sr-90 analysis by accelerator mass spectrometry
加速器質量分析によるSr-90の高感度法
笹 公和*; 本多 真紀
; 細谷 青児*; 高橋 努*; 高野 健太*; 落合 悠太*; 坂口 綾*; 栗田 沙緒里*; 佐藤 志彦
; 末木 啓介*
Sasa, Kimikazu*; Honda, Maki; Hosoya, Seiji*; Takahashi, Tsutomu*; Takano, Kenta*; Ochiai, Yuta*; Sakaguchi, Aya*; Kurita, Saori*; Satou, Yukihiko; Sueki, Keisuke*
Strontium-90 (
Sr) is one of the most important fission products due to the potential health risks of its uptake and retention in the human body. Conventional analysis techniques involve beta-counting, which requires ingrowth of
Y over a period of two weeks or more. Accelerator mass spectrometry (AMS) has the potential to shorten the analysis time while offering a lower limit of detection than beta-counting. Here, Sr in samples was recovered as SrF
to provide sufficient negative ions in the caesium-sputtering ion source. In the sample preparation step, 95-98% of Sr was recovered and 99-100% of Zr removed by ion-exchange separation. Sr recovery was
30% in the precipitation process, and this can be improved. The recovered
Sr was mixed with PbF
at an SrF
:PbF
weight ratio of 4:1. A maximum 500 nA beam current of Sr
ions was obtained from SrF
samples. A five-anode gas ionization detector was used to avoid isobaric interference from
Zr. The
Sr/Sr atomic ratio background of
6
10
(equivalent to
3 mBq
Sr) was comparable with that achieved at other AMS facilities. Good linearity
Sr/Sr atomic ratios was obtained from 1.75
10
to 3.38
10
. Suitable techniques for sample preparation and measurement were thus achieved for
Sr analysis by AMS.